Отрывок: To acquire images for calibration and tests, we utilized two types of plane calibration targets: (1) with black circular dots on a white background and (2) with black and white chessboard pattern. Due to sufficient range of distances and magnifications, three samples of each calibration target type with 0.5 mm (small-sized), 1 mm (middle-sized) and 2 mm (large-sized) distance between markers have been manufac- tured. This distance corr...
Полная запись метаданных
Поле DC Значение Язык
dc.contributor.authorGorevoy, A.V.-
dc.contributor.authorMachikhin, A.S.-
dc.date.accessioned2017-10-25 12:06:09-
dc.date.available2017-10-25 12:06:09-
dc.date.issued2017-08-
dc.identifierDspace\SGAU\20171020\65756ru
dc.identifier.citationGorevoy AV, Machikhin AS. Optimal calibration of a prism-based videoendoscopic system for precise 3D measurements. Computer Optics 2017; 41(4): 535-544.ru
dc.identifier.urihttps://dx.doi.org/10.18287/2412-6179-2017-41-4-535-544-
dc.identifier.urihttp://repo.ssau.ru/handle/Zhurnal-Komputernaya-optika/Optimal-calibration-of-a-prismbased-videoendoscopic-system-for-precise-3D-measurements-65756-
dc.description.abstractModern videoendoscopes are capable of performing precise three-dimensional (3D) measurements of hard-to-reach elements. An attachable prism-based stereo adapter allows one to register images from two different viewpoints using a single sensor and apply stereoscopic methods. The key condition for achieving high measurement accuracy is the optimal choice of a mathematical model for calibration and 3D reconstruction procedures. In this paper, the conventional pinhole camera models with polynomial distortion approximation were analyzed and compared to the ray tracing model based on the vector form of Snell’s law. We, first, conducted a series of experiments using an industrial videoendoscope and utilized the criteria based on the measurement error of a segment length to evaluate the mathematical models considered. The experimental results confirmed a theoretical conclusion that the ray tracing model outperforms the pinhole models in a wide range of working distances. The results may be useful for the development of new stereoscopic measurement tools and algorithms for remote visual inspection in industrial and medical applications.ru
dc.description.sponsorshipThe Russian Science Foundation (project #7-19-01355) financially supported the work. The authors thank A. Naumov, A. Shurygin and D. Khokhlov for continuous technical support.ru
dc.language.isoenru
dc.publisherСамарский университетru
dc.relation.ispartofseries41;4-
dc.subjectremote visual inspectionru
dc.subject3D spatial measurementsru
dc.subjectcamera calibrationru
dc.subjectdistortion modelru
dc.subjectprism distortionru
dc.titleOptimal calibration of a prism-based videoendoscopic system for precise 3D measurementsru
dc.typeArticleru
dc.textpartTo acquire images for calibration and tests, we utilized two types of plane calibration targets: (1) with black circular dots on a white background and (2) with black and white chessboard pattern. Due to sufficient range of distances and magnifications, three samples of each calibration target type with 0.5 mm (small-sized), 1 mm (middle-sized) and 2 mm (large-sized) distance between markers have been manufac- tured. This distance corr...-